DL-Warner 6126B

Automess

DL-Warner 6126B

The DL-WARNER 6126B is a battery-operated dose rate alarm for photon radiation (gamma and X-rays) expressed in the unit Hp(10). The device is designed to provide personal protection when handling radioactive materials or whilst in areas at risk of radiation exposure. It can be worn comfortably on one’s clothing.

An energy-compensated Geiger–Müller counter is used as the radiation detector. The device is characterised by its ease of use and robust construction.

The control switch is located on the upper narrow side of the device, making it easily accessible. The switch has three positions: ‘Off’, ‘Batt.’ and ‘On’. A further switch allows the warning threshold to be set to three different dose rate values. The shaft of this switch has a slot, so a screwdriver is required to adjust it. This prevents the warning threshold from being adjusted accidentally.

The ‘Batt.’ switch position is used to check the battery and the device’s functionality. In this position, the alarm tone indicates that the battery still guarantees a minimum operating time of 30 hours. The battery compartment is located under a sliding cover in the lower part of the device.

If, in the ‘On’ position, the dose rate at the device’s location exceeds the set threshold, a clearly audible alarm sounds until the dose rate falls below the threshold or the device is switched off.

  • Measured quantity: Deep personal dose Hp(10)
  • Adjustable warning thresholds: 25 µSv/h (optionally 10 µSv/h), 1 mSv/h, 10 mSv/h
  • Energy range: 50 keV – 1.3 MeV
  • Fire service approval: DLW/FW/IdF 027201/4 including
    9V battery (IEC 6LR61)